myLabComplianceCLIA compliance software · lab-consultants.comBook a demo

Proficiency testing tracking for CLIA laboratories

What CLIA requires for proficiency testing (PT), the mistakes that cause serious findings, and how to track every event to closure.

Book a demo Visit mylabcompliance.io

What proficiency testing is

Proficiency testing is external quality assessment: an approved provider sends unknown samples, the lab tests them like patient specimens, and the provider grades the results against peer or reference values. CLIA's PT requirements are in Subpart H of 42 CFR Part 493 (eCFR), beginning at 42 CFR 493.801. Program and grading criteria are in Subpart I.

Core CLIA requirements

  • Enrollment: labs performing nonwaived testing must enroll in a CMS-approved PT program for each specialty, subspecialty and regulated analyte they test, and notify the program which results CMS may see (42 CFR 493.801(a)).
  • Testing like patients: PT samples are tested with the regular patient workload, by personnel who routinely test patient samples, using the lab's routine methods (42 CFR 493.801(b)(1)).
  • Attestation: the testing personnel and the laboratory director sign an attestation that samples were handled routinely (42 CFR 493.801(b)(1)).
  • No communication: labs must not communicate with other labs about PT results before the event's submission deadline (42 CFR 493.801(b)(3)).
  • No referral: labs must not send PT samples or portions of them to another lab for testing, even if they normally refer patient specimens for that test (42 CFR 493.801(b)(4)). Intentional referral can lead to revocation of the CLIA certificate under 42 CFR 493.1840.
  • Records: keep copies of all records, including worksheets and attestations, for the period required (42 CFR 493.801(b)(5)).

Scoring: satisfactory, unsatisfactory and unsuccessful

PT providers typically send three events a year. For most regulated analytes, a score of at least 80% for an event is satisfactory; some areas, such as ABO group and D typing and compatibility testing, require higher scores (see Subpart I). An unsatisfactory event score should trigger investigation and documented corrective action. Unsuccessful participation, defined in 42 CFR 493.2 as unsatisfactory performance in two consecutive or two out of three testing events, can lead to CMS sanctions, including suspending testing of that analyte until the lab demonstrates sustained satisfactory performance (42 CFR 493.803).

A failure to participate or to return results on time is scored as zero. Missing a deadline is as damaging as a wrong answer.

Tests without required PT

For tests not on the regulated analyte list, CLIA still requires the lab to verify accuracy at least twice a year (42 CFR 493.1236(c)). Options include enrolling in a PT program anyway, split-sample comparison with another lab, or other documented approaches. These must be tracked just like regulated PT.

A PT tracking workflow

  1. Build the enrollment map: every test on your menu, whether it is regulated, which program and module covers it, or which alternative assessment applies.
  2. Load the calendar: shipment dates, testing windows and submission deadlines for each event.
  3. Assign testing personnel and rotate among staff who perform patient testing.
  4. Test and attest: record results, worksheets and signed attestations.
  5. Review evaluations: the director or designee reviews each graded report and signs.
  6. Investigate failures: for each unacceptable result, check clerical error, methodology, equipment, QC, calibration and personnel; assess whether patient results were affected; document corrective action.
  7. Trend: look for repeated near-misses or bias even on passing events.

Common PT findings

  • No documented review of PT evaluations.
  • Unacceptable results with no investigation, or an investigation that stops at "clerical error" without correction.
  • PT always run by the same senior tech.
  • Missing attestations.
  • No twice-yearly accuracy verification for unregulated tests.
  • PT samples sent to a reference lab because "that's where we send patient samples," which is prohibited.

How myLabCompliance tracks PT

myLabCompliance lists PT enrollment tracking, result submission management, performance analysis, corrective action documentation, a PT calendar and reminders among its features. PT records sit alongside QC, competency and instrument records, which matters because PT results can also serve as evidence for competency assessment and because investigations often lead back to QC or calibration.

Related: competency tracking, quality management, CLIA checklist.

Frequently asked questions

What score is needed to pass a CLIA PT event?

For most regulated analytes, at least 80% is satisfactory. Some areas, such as ABO/D typing and compatibility testing, require higher scores, as set out in 42 CFR Part 493 Subpart I.

What is unsuccessful PT participation?

Unsatisfactory performance in two consecutive events or two of three consecutive events for the same analyte, specialty or subspecialty, as defined in 42 CFR 493.2.

Can we send PT samples to our reference lab?

No. 42 CFR 493.801(b)(4) prohibits referring PT samples to another laboratory, even if you refer patient specimens for that test.

What about tests that have no PT requirement?

42 CFR 493.1236(c) requires verifying the accuracy of those tests at least twice a year.

Can PT results be used in competency assessment?

Yes. CLIA's competency procedures include assessing test performance through testing previously analyzed specimens, internal blind samples or external PT samples.

See myLabCompliance with your own lab's workflow

Book a demo and bring the compliance task that takes the most of your week. Product details are on mylabcompliance.io.

Book a demo

Sources: 42 CFR 493.801 · 42 CFR 493.803 · 42 CFR 493.1236 · 42 CFR 493.2 · 42 CFR Part 493 (eCFR) · CMS CLIA program

This page is general information about laboratory regulations, not legal or regulatory advice. Confirm current requirements with CMS, your state agency, and your accrediting organization.